Section 9: Verifying the Model 9500B Accuracy Specification 9-31
Final Width = 215mm
*: Verification points EDG11 to EDG20 available for Active Head Model 9530 only.
†: 9500B displayed value accuracy specification: for
500ps
, ±40ps; for
150ps
, ±15ps; for , ±11ps; for , ±4ps.
EDG01
500ps
25.000mV -150ps +50ps ps ps ps ps
EDG02 ps ps ps ps
EDG03
500ps
50.000mV -150ps +50ps ps ps ps ps
EDG04 ps ps ps ps
EDG05
500ps
250.00mV -150ps +50ps ps ps ps ps
EDG06 ps ps ps ps
EDG07
500ps
500.00mV -150ps +50ps ps ps ps ps
EDG08 ps ps ps ps
EDG09
500ps
2.5000V -150ps +50ps ps ps ps ps
EDG10 ps ps ps ps
EDG11
150ps
* 25.000mV -25ps +25ps ps ps ps ps
EDG12 ps ps ps ps
EDG13
150ps
* 50.000mV -25ps +25ps ps ps ps ps
EDG14 ps ps ps ps
EDG15
150ps
* 250.00mV -25ps +25ps ps ps ps ps
EDG16 ps ps ps ps
EDG17
150ps
* 500.00mV -25ps +25ps ps ps ps ps
EDG18 ps ps ps ps
EDG19
150ps
* 2.5000V -25ps +25ps ps ps ps ps
EDG20 ps ps ps ps
Table 9.9.3.1 — Edge Function Verification — Rise/Fall Time at 1MHz (Period 1µs) into 50 W Load
Please copy the following table. Enter the calculation results and measurements in the appropriate columns on the copy:
9.9.3 Verifying the Edge Function (Contd.)
70ps 25ps
Ver. Nominal Edge Output Lower Upper User's Lower Upper Measured 9500B
Pt. Rise/Fall Type Voltage Tolerance Tolerance Total Validity Validity Value Displayed
Time Limit Limit Measurement Tolerance Tolerance 'Rise Time'
(10% - 90%) (pk-pk) Uncertainty Limit Limit Value

Table of Contents