Infratec™ 1241 Grain Analyzer
Service Manual 1001 5015 / Rev. 4 4:53
SELFTEST/12417132/Monochromator/SerialNumber=String,E0113281
SELFTEST/12417132/Monochromator/PulseCount=Number,2884 (>2800 , <3400)
SELFTEST/12417132/Monochromator/CycleTime=Number,2.74 (>2.0 , <3.2)
SELFTEST/12417132/Monochromator/Min=Number,-4.7 (>-22)
SELFTEST/12417132/Monochromator/Max=Number,5.6 (<22)
SELFTEST/12417132/Monochromator/Diff=Number,10.3 (<50)
SELFTEST/12417132/END MONOCHROMATOR TEST/2011-01-18 15:42:56
SELFTEST/12417132/START DETECTOR TEST/2011-01-18 15:43:03
SELFTEST/12417132/Detector/SerialNumber=String,001601
SELFTEST/12417132/Detector/Gain1/Min=Number,688 (>=5)
SELFTEST/12417132/Detector/Gain1/Max=Number,689
SELFTEST/12417132/Detector/Gain1/Mean=Number,688.3450 (>50 , <2000)
SELFTEST/12417132/Detector/Gain1/StdDev=Number,0.4754 (<= 2)
SELFTEST/12417132/Detector/Gain1/Temp=Number,34.0 (>25.0 , <65.0)
SELFTEST/12417132/2011-01-18 15:43:03
SELFTEST/12417132/Detector/Gain2/Min=Number,687 (>=5)
SELFTEST/12417132/Detector/Gain2/Max=Number,688
SELFTEST/12417132/Detector/Gain2/Mean=Number,687.4498 (>50 , <2000)
SELFTEST/12417132/Detector/Gain2/StdDev=Number,0.4975 (<= 2)
SELFTEST/12417132/Detector/Gain2/Temp=Number,34.0 (>25.0 , <65.0)
SELFTEST/12417132/2011-01-18 15:43:03
SELFTEST/12417132/Detector/Gain3/Min=Number,685 (>=5)
SELFTEST/12417132/Detector/Gain3/Max=Number,686
SELFTEST/12417132/Detector/Gain3/Mean=Number,685.7570 (>50 , <2000)
SELFTEST/12417132/Detector/Gain3/StdDev=Number,0.4289 (<= 2)
SELFTEST/12417132/Detector/Gain3/Temp=Number,34.0 (>25.0 , <65.0)
SELFTEST/12417132/2011-01-18 15:43:04
SELFTEST/12417132/Detector/Gain4/Min=Number,681 (>=5)
SELFTEST/12417132/Detector/Gain4/Max=Number,683
SELFTEST/12417132/Detector/Gain4/Mean=Number,681.8250 (>50 , <2000)
SELFTEST/12417132/Detector/Gain4/StdDev=Number,0.3883 (<= 2)
SELFTEST/12417132/Detector/Gain4/Temp=Number,34.0 (>25.0 , <65.0)
SELFTEST/12417132/2011-01-18 15:43:04