Infratec™ 1241 Grain Analyzer
4:54 Service Manual 1001 5015 / Rev. 4
SELFTEST/12417132/Detector/Gain5/Min=Number,673 (>=5)
SELFTEST/12417132/Detector/Gain5/Max=Number,675
SELFTEST/12417132/Detector/Gain5/Mean=Number,674.0966 (>50 , <2000)
SELFTEST/12417132/Detector/Gain5/StdDev=Number,0.4810 (<= 2)
SELFTEST/12417132/Detector/Gain5/Temp=Number,34.0 (>25.0 , <65.0)
SELFTEST/12417132/2011-01-18 15:43:04
SELFTEST/12417132/Detector/Gain6/Min=Number,689 (>=5)
SELFTEST/12417132/Detector/Gain6/Max=Number,690
SELFTEST/12417132/Detector/Gain6/Mean=Number,689.4406 (>50 , <2000)
SELFTEST/12417132/Detector/Gain6/StdDev=Number,0.4965 (<= 2)
SELFTEST/12417132/Detector/Gain6/Temp=Number,34.0 (>25.0 , <65.0)
SELFTEST/12417132/2011-01-18 15:43:05
SELFTEST/12417132/Detector/Gain7/Min=Number,689 (>=5)
SELFTEST/12417132/Detector/Gain7/Max=Number,690
SELFTEST/12417132/Detector/Gain7/Mean=Number,689.4748 (>50 , <2000)
SELFTEST/12417132/Detector/Gain7/StdDev=Number,0.4994 (<= 2)
SELFTEST/12417132/Detector/Gain7/Temp=Number,34.0 (>25.0 , <65.0)
SELFTEST/12417132/2011-01-18 15:43:05
SELFTEST/12417132/Detector/Gain8/Min=Number,688 (>=5)
SELFTEST/12417132/Detector/Gain8/Max=Number,690
SELFTEST/12417132/Detector/Gain8/Mean=Number,689.3686 (>50 , <2000)
SELFTEST/12417132/Detector/Gain8/StdDev=Number,0.5099 (<= 3)
SELFTEST/12417132/Detector/Gain8/Temp=Number,34.0 (>25.0 , <65.0)
SELFTEST/12417132/2011-01-18 15:43:05
SELFTEST/12417132/Detector/Gain9/Min=Number,687 (>=5)
SELFTEST/12417132/Detector/Gain9/Max=Number,691
SELFTEST/12417132/Detector/Gain9/Mean=Number,689.1770 (>50 , <2000)
SELFTEST/12417132/Detector/Gain9/StdDev=Number,0.8731 (<= 3)
SELFTEST/12417132/Detector/Gain9/Temp=Number,34.0 (>25.0 , <65.0)
SELFTEST/12417132/2011-01-18 15:43:06
SELFTEST/12417132/Detector/Gain10/Min=Number,684 (>=5)
SELFTEST/12417132/Detector/Gain10/Max=Number,694
SELFTEST/12417132/Detector/Gain10/Mean=Number,689.0166 (>50 , <2000)
SELFTEST/12417132/Detector/Gain10/StdDev=Number,1.5825 (<= 4)