D-6 250cx Series Maternal/Fetal Monitor Revision C
2036946-001
Frequently Asked Questions:
How do I change the alarm limits for
MSpO
2
, Maternal Blood Oxygen
Saturation?
1. Rotate the Trim Knob to highlight the legend for MSpO
2
. (This legend is slightly
above center, on the right side of the display.)
2. Once the MSpO
2
legend is highlighted, press the Trim Knob. The display changes to
show the MSpO
2
Setup screen.
3. Now rotate the Trim Knob to highlight the current High: saturation percentage (%)
alarm limit setting.
4. Once the High: saturation percentage (%) alarm limit setting is highlighted, press the
Trim Knob again. The current setting displays in blinking inverse video.
5. Now rotate the Trim Knob to change the current setting. Select from 100 to 85%.
6. Repeat 3 through 6 for the High: saturation percentage (%) alarm limit setting. Select
a value between 99% and 80%.
You may optionally change the MHR/P (Maternal Heart Rate/Pulse) High: select a
value between 100 bpm and 250 bpm. (The MHR/P alarm settings are also
available through the Pulse legend, or the NIBP legend on the main monitoring
screen.)
You may optionally change the MHR/P Low: select a value between 35 bpm and
120 bpm. (The MHR/P alarm settings are also available via the Pulse legend, or
the NIBP legend on the main monitoring screen.)
7. Once the desired source parameter is set, press the Trim Knob to confirm your
selection. The current value setting stops blinking.
8. Now rotate the Trim Knob to select (highlight) the Exit item on the bottom menu.
9. Press the Trim Knob again to return to the main monitoring display screen.
How do I change the waveform parameter
being displayed?
1. Rotate the Trim Knob to highlight the waveform selection item. (This legend is
slightly above the bottom menu bar, to the far left of the display, and should indicate
one of the following: Off, MECG, MSpO
2
, or FECG.)
2. Once the waveform selection item is highlighted, press the Trim Knob again. The
current setting displays in blinking inverse video.
3. Now rotate the Trim Knob to change the setting to Off, MECG, MSpO
2
, or FECG.
4. Once you set the desired source parameter, press the Trim Knob to confirm your
selection. The current value setting stops blinking.
5. The selected waveform parameter displays on the main monitoring screen.
Question Answer

Table of Contents

Other manuals for GE Corometrics 250cx Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GE Corometrics 250cx Series and is the answer not in the manual?

GE Corometrics 250cx Series Specifications

General IconGeneral
TypeFetal Monitor
Model250cx Series
ManufacturerGE Healthcare
Fetal Heart Rate MonitoringYes
Ultrasound Frequency1 MHz
Recorder Speed1, 2, 3 cm/min
Maternal ParametersECG, NIBP, SpO2
FHR Range30-240 bpm
TOCO Range0-100
ConnectivityRS-232
Power SourceAC Power
DisplayColor LCD

Related product manuals