D-8 250cx Series Maternal/Fetal Monitor Revision C
2036946-001
Frequently Asked Questions:
How do I disable/enable Fetal Movement
Detection? (FMD is an optional feature that
is purchased separately.)
1. Rotate the Trim Knob to highlight the legend for US. (This legend is top left on the
display.)
Note: This feature applies only if ultrasound is the source.
2. Once the US legend is highlighted, press the Trim Knob. The display changes to
show the US Setup screen where US (not US2 Setup) is the current legend.
3. Now rotate the Trim Knob to highlight the FM Detect: setting. This setting is On or
Off.
4. Once the FM Detect: setting is highlighted, press the Trim Knob again. The current
setting displays in blinking inverse video.
5. Now rotate the Trim Knob to change the current setting. Select a value of On
(enabled) or Off (disabled).
6. Once you set the desired value, press the Trim Knob again to save the value. The
current value setting stops blinking.
7. Now rotate the Trim Knob to select (highlight) the Exit item on the bottom menu.
8. Press the Trim Knob again to return to the main monitoring display screen.
How do I enable and change the alarm
volume settings for Fetal Heart Rate 1?
1. Rotate the Trim Knob to highlight the legend for FHR1. (This legend is top left on the
display, and it may read INOP, FECG, US, or US2.)
2. Once the FHR1 legend is highlighted, press the Trim Knob. The display changes to
show the <MODE> Setup screen, where <MODE> is the current legend.
3. Now rotate the Trim Knob to highlight the FHR1 Audio Alarms: setting. This is at the
bottom of the flat panel display, slightly above the menu bar, on the left half of the
screen. The setting is either On or Off. If the setting is Off, it needs to be enabled.
Proceed to (Step 4) below. If it is already On, and you simply wish to change the
Volume: setting, proceed to (Step 7) below.
4. Once the Audio Alarms: setting is highlighted, press the Trim Knob again. The
current setting displays in blinking inverse video.
5. Now rotate the Trim Knob to change the current setting to On.
6. Once you set the desired alarm value, press the Trim Knob to confirm your selection.
The current value setting stops blinking.
7. Now rotate the Trim Knob to select (highlight) the Volume: setting which is located
immediately to the right. The setting is in the range of 1 to 9.
8. Once the Volume: setting is highlighted, press the Trim Knob again. The current
setting displays in blinking inverse video, and an audio tone, indicative of the alarm
volume, is emitted from the speaker.
9. Now rotate the Trim Knob to change the current setting, as desired. Each time you
change a setting, you will hear an audio tone, indicative of the alarm volume, emitted
from the speaker.
10. Once you set the desired alarm value, press the Trim Knob to confirm your
selection. The current value setting stops blinking. A final audio tone, indicative of the
alarm volume, is emitted from the speaker.
11. Now rotate the Trim Knob to select (highlight) the Exit item on the bottom menu.
12. Press the Trim Knob again to return to the main monitoring display screen.
Question Answer

Table of Contents

Other manuals for GE Corometrics 250cx Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GE Corometrics 250cx Series and is the answer not in the manual?

GE Corometrics 250cx Series Specifications

General IconGeneral
TypeFetal Monitor
Model250cx Series
ManufacturerGE Healthcare
Fetal Heart Rate MonitoringYes
Ultrasound Frequency1 MHz
Recorder Speed1, 2, 3 cm/min
Maternal ParametersECG, NIBP, SpO2
FHR Range30-240 bpm
TOCO Range0-100
ConnectivityRS-232
Power SourceAC Power
DisplayColor LCD

Related product manuals