3-18 250cx Series Maternal/Fetal Monitor Revision C
2036946-001
Controls, Indicators, and Connectors: Rear Panel Description
CAUTION
NON-DESTRUCTIVE VOLTAGE—The maximum non-
destructive voltage that may be applied to the rear panel
connectors is 0 volts. Do not attempt to connect cables to these
connectors without contacting your Biomedical Engineering
Department or GE Medical Systems Information Technologies
Service Representative. This is to ensure the connectors comply
with leakage-current requirements of one of the following
applicable standards: Underwriters Laboratories UL-2601-1,
Canadian Standards Associations CSA 22.2 No. 125, or
International Electrotechnical Commission EN60601-1.
M
AC Voltage Selection
Switch
This switch is intended for qualified service
personnel to select a voltage range for the AC
input:
120: Accepts an AC input in the range of 100–
120 VAC.
240: Accepts an AC input in the range of 220–
240 VAC
N
Power Entry Module AC line power cord receptacle. Refer to the rear
panel markings to verify line voltage and line
frequency requirements.
250cx Series Rear Panel (Standard and Optional Features)
Name Description

Table of Contents

Other manuals for GE Corometrics 250cx Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GE Corometrics 250cx Series and is the answer not in the manual?

GE Corometrics 250cx Series Specifications

General IconGeneral
TypeFetal Monitor
Model250cx Series
ManufacturerGE Healthcare
Fetal Heart Rate MonitoringYes
Ultrasound Frequency1 MHz
Recorder Speed1, 2, 3 cm/min
Maternal ParametersECG, NIBP, SpO2
FHR Range30-240 bpm
TOCO Range0-100
ConnectivityRS-232
Power SourceAC Power
DisplayColor LCD

Related product manuals