•  In the H201Ti:  To the Supervisory Link and Analog Output Termination Board 
(item 9 in Figure 3-4). 
•  In the H201Ci-C:  To the terminals 21 to 24 of the Main Terminal Block (item 5 
in Figure 4-2). 
•  In the H201Ci-1:  To the terminals 21 to 24 of the Main Terminal Block (item 6 
in Figure 4-6). 
10.1.1 TDM Pulse 
The TDM pulse is used only with the Hydran 201Ci-1 Controller to transfer two pieces of 
information from the Hydran 201Ti to the H201Ci-1 itself: 
•  Actual  dissolved  gas  level:    This  analog  signal  varies  from  0  to  100%  (0  to 
2,000ppm) with a resolution of 1 part in 4,000. 
•  Alarm relays status:  This digital information has four possible values, namely 
Normal, High, High-High and Fail. 
The relationship between the alarm relay states and the status of the H201Ti, of the TDM 
signal and of the H201Ci-1 is shown in Table 10-1 below. 
a.  - = without consequence; H = High Alarm; HH = High-High Alarm; Sup = Supervisory. 
b.  On = Normal (no alarm); Off =  Alarm 
Table 10-1: Alarm Relay States and Status of the H201Ti, TDM Signal and H201Ci-1