11.5.3 Fail Alarm Cancellation 
Note:    To  be  cancelled,  an  alarm  condition  must  disappear  during  a  period  of  time 
exceeding the corresponding alarm delay (if existing for this condition). 
To cancel a fail alarm, proceed as follows: 
1.  Find and solve the problem (for help, see Section 11.5.2). 
2.  Acknowledge  the  alarm  using  the  Hydran  201Ti’s  keypad  or  the  Hydran  Host 
software.    For  the  procedure,  see  Section  5.2.1,  and  the  Hydran  Host  Software 
Manual.  If the problem has been corrected during a period of time exceeding the 
corresponding  alarm  delay,  the  alarm  message  then  disappears  from  the 
H201Ti’s display; otherwise, it simply stops flashing and remains on the display.