9.2.4.6  Conclusion 
If necessary, set the parameter RelayMode of each alarm relay to their initial mode. 
 
9.2.5  Verifying the RS 232 or USB Communications 
If your installation includes only one Hydran 201R Model i and if it communicates with 
the host computer running the Hydran Host or Perception software, you must verify the 
RS-232 link or USB link depending which one is used. Proceed as follows: 
1.  Connect  the  Hydran  201Ci-1’s  RS-232  Computer  Port  connector  (J10;  item  7  in 
Figure 4-7) to your PC’s RS-232 port.  For wiring details, see Table D - 13.  For USB 
users, connect the Hydran 201Ci-1’s USB Computer Port connector (J16; item 8 in 
Figure 4-7) to your PC’s USB port. 
2.  In  the  Communication  submenu  of  the  Hydran  201Ti,  write  down  the  current 
settings, in order to be able to restore them. 
3.  Set the parameters as follows: 
•  Comm Channel = Supervisory Link  
•  Default Channel = Supervisory Link  
•  Baud Rate = 9600  
•  Gas Monitor ID = 1  
 
4.  To establish the communication and depending on the protocol configured in the 
201Ti  devices, start the  Hydran Host  or  Perception software  on  the PC.  See  the 
Hydran Host Software Manual or Perception User Manual for more information. 
5.  If the communication cannot be established:  
•  Ensure the H201Ti identification number entered in the Hydran Host software 
corresponds to the one entered in the H201Ti (Gas Monitor ID parameter). 
•  Verify the wiring of the supervisory link cable. 
•  Verify  the  configuration  of  the  Hydran  Host  software.    See  the  Hydran  Host 
Software Manual or Perception User Manual for more information.  
 
6.  Disconnect all communication cables (supervisory link and RS-232 or USB to the 
PC).  
7.  When  the  test  is  completed,  restore  the  original  communication  settings  in  the 
H201Ti as noted in the step 2 above.  
 
9.2.6  Conclusion 
This completes the incoming verification of the Hydran 201Ci-1. 
1.  Disconnect the H201Ti from the power outlet. 
2.  Disconnect the H201Ci-1 from the power outlet.