8.1.3  Storage 
If  the  Hydran  201Ci-C  is  not  installed  immediately,  it  is  strongly  recommended  to 
store it in its original shipping box in a safe place. 
Storage should be at a temperature between 5 and 45°C (41 and 113°F) and a non-
condensing,  relative  humidity  between  5  and  95%;  if  these  limits  are  to  be  exceeded, 
please contact the General Electric Customer Service (contact information can be found 
at the bottom of page 2). These restrictions do not apply to transportation. 
 
8.2  Verification of Operation Prior to Installation 
It is strongly recommended to verify the operation of the Hydran 201Ci-C Controller(s) 
before installation in the field.  The verification confirms that no damage occurred during 
shipping and allows the user to become familiar with the H201Ci-C. 
 
8.2.1  Preparation 
To verify its operation,  the Hydran  201Ci-C must first be connected to  at least one 
functional Hydran  201Ti via  a  supervisory link  cable.   One  end of this cable  is  already 
connected to the H201Ti.  For details on commissioning the H201Ti(s), see Chapter 7. 
Proceed as follows: 
1.  Connect  a  three-wire power cord to the AC Power 3-Terminal  Block on  the first 
H201Ti (item 8 in Figure 3-4):  The live side to L1/L, the other side to L2/N, and the 
ground to E/G.  For additional wiring details, see Table D - 1. 
Disconnect  power  at  disconnect  switch  or  circuit  breaker 
before making any electrical connections, and ensure a proper 
ground connection is made before connecting line voltage. 
Failure to do so can result in property damage, personal injury 
and/or death. 
2.  Connect the second end of the two-pair supervisory link cable coming out of the 
H201Ti to the main terminal block on the H201Ci-C (item 5 in Figure 4-2). 
3.  Power On the H201Ti. 
4.  Ensure the H201Ti is operating normally and no gas or fail alarms are present, by 
looking at the small  LED indicators on the right side of the I/O  module (refer to 
Section 3.4.6.3). 
5.  Repeat the steps 1 to 4 above until all H201Tis are connected to the H201Ci-C. 
6.  Connect  a  power cord to  the  AC  Supply Connector on the H201Ci-C  (item 19 in 
Figure 4-3). 
7.  Power On the H201Ci-C. 
8.  Repeat the steps 1 to 7 above until all H201Ci-C’s are fed.