15  DECOMMISSIONING THE HYDRAN 201TI 
The  Hydran  201i  system  is  designed  to  be  permanently  installed  for  autonomous 
operation on a transformer. If however, the Hydran 201i must be decommissioned, the 
procedure to be followed is described in this chapter. 
15.1  Powering Off 
The  following  procedure  must  be  carried  out  thoroughly. 
Failure to do so may result in property damage, personal injury 
and/or death. 
The decommissioning of the Hydran 201Ti must be executed by 
trained service personnel only. 
1.  Switch  OFF  the  circuit  breaker  to  which  the  power  cable  of  the  Hydran  201i 
system is connected. 
2.  Verify, with a voltmeter, there is no AC or DC voltage on the power cable. 
3.  Disconnect the power cable from the circuit breaker. 
4.  Remove the cover of the Hydran 201i system to be decommissioned. 
5.  Verify there is no AC or DC voltage on the mains supply terminals of the unit. 
6.  Verify there is no AC or DC voltage on the cables connected to the analog inputs, 
analog outputs or/and digital inputs if present. 
7.  Verify there is no AC or DC voltage on the cables connected to the alarm relays. 
8.  Verify  there  is  no  AC  or  DC  voltage  on  the  cable  connected  to  the  RS-485 
terminals. 
9.  If  any  of  the  cables  is  still  fed  by  an  external  source,  identify  and  isolate  the 
source. 
15.2  Cable Disconnection 
After verification that there is no AC or DC voltage on all cables connected to the Hydran 
201Ti unit, proceed as following: 
1.  Disconnect the power supply cable and remove this cable. 
2.  Disconnect  the  cables  from  the  analog  inputs,  analog  outputs  or/and  digital 
inputs if installed and remove these cables. 
3.  Disconnect the cable from the alarm relays terminals and remove this cable. 
4.  Disconnect the cable from the RS-485 terminals and remove this cable.