14.5  Historic Data Verification 
Frequency:  Annually or on alarm 
In  addition  to  the  annual or  on-alarm  verification,  the  Hydran  201Ti’s  historic  data 
must also be read regularly by the user. 
If the data is read from the H201Ti, follow the instructions in Section 5.5.1. The data 
can  be  collected  using  a  host  computer  running  the  Hydran  Host  software  (see  the 
Hydran Host Software Manual). 
The historic data are verified as follows: 
1.  Consult  the  historic  data  file  Events  to  investigate  suspicious  alarms  and  other 
events.  
2.  Consult the historic data files Short Term and Long Term to study the evolution of 
the gas level and the hourly and daily trends.  
3.  Ensure the hourly  and daily trends  periods  are optimized.   Change the values if 
necessary.  
4.  Examine every alarm setting.  Change the values if necessary.  
Note:  Alarm settings should be modified by authorized personnel only. 
 
14.6  Dissolved Gas Analysis (DGA) 
Frequency:  Annually (minimum) or on alarm 
The dissolved gas analysis (DGA) is the reference method used to determine the exact 
level of dissolved gases in oil. A DGA should be performed at least once a year. 
To verify the readings of the Hydran 201i System using a DGA, proceed as follows: 
1.  Note the Hydran 201Ti’s gas level reading. 
2.  Take an oil sample from the Hydran 201 sensor’s sampling port (see Appendix I).  
3.  Send the sample to a qualified laboratory. 
The DGA results include the concentrations, in parts per million (ppm), of the following 
gases: 
•  Hydrogen (H
2
)  
•  Carbon monoxide (CO)  
•  Acetylene (C
2
H
2
) 
•  Ethylene (C
2
H
4
) 
•  Methane (CH
4
) 
•  Ethane (C
2
H
6
) 
•  Carbon dioxide (CO
2
) 
•  Nitrogen (N
2
) 
•  Oxygen (O
2
)