12.1  Operating Methods 
The  Hydran  201  i  System  can  be  operated  using  several  methods.    Each  of  these 
methods is explained in the following pages, from the simplest to the most powerful one: 
•  Periodic, visual monitoring:  See Section 12.1.1below. 
•  Alarm monitoring:  See Section 12.1.2. 
•  Analog output monitoring:  See Section 12.1.3. 
•  Combined alarm and analog output monitoring:  See Section 12.1.4. 
•  Local monitoring with host computer:  See Section 12.1.5. 
•  Remote monitoring with host computer:  See Section 12.1.6. 
 
12.1.1 Periodic, Visual Monitoring 
Among the Hydran 201i System’s operating methods, the periodic, visual monitoring 
is  the  simplest.    If  used,  the  following  routine  must  be  an  integral  part  of  the  on-site, 
station inspection: 
•  Periodic, visual reading (once or twice a day) of the gas level reading on the 
display of the Hydran 201Ti’s or the Hydran 201Ci-1.  For details on the various 
methods of data reading, see Section 12.2. 
•  On-site verification of alarms.  For details on the alarms and relays operation, 
see Chapter 11. 
•  Manual logging of the above results. 
Only  basic  training  to  the  station’s  personnel  is  required  to  implant  this  operating 
method.  When an alarm is detected, the operator must report the alarm and take the 
appropriate  actions  for  this  type  of  alarm  (follow  the  regulations  and  practices  of  the 
company). 
Note:  With this method, the analog outputs and the alarm contacts are not connected to 
a SCADA system. 
1.1.1.12 Method Drawbacks 
The periodic,  visual  monitoring  method  has a  deficient response  time  following  the 
detection  of  an  alarm;  this  response  time  cannot  be  inferior  to  the  period  of  time 
between  two  inspections.    The  Hydran  201i  System,  however,  has  a  response  time 
measured in minutes.