GE HEALTHCARE
DIRECTION 2286865, REVISION 14 LOGIQ™ 7 SERVICE MANUAL
Section 8-2 - Software Loading Procedures for BT09 (R8.x.x or later) 8-45
8-2-10-7 Regional Setup (For the system with language other than English ONLY) (cont’d)
4.) Select the Advanced tab, then select Russian or Greek in the Language for non-Unicode
programs pull-down menu. Press Apply. Answer Yes to use files already loaded on the hard disk,
then answer No to not reboot the system yet, press OK. Press Save and Exit the Utility screen.
5.) Reboot the system. When your system restarts, the system appears in the selected language.
6.) To switch between the English and Russian (or Greek) keyboard, press Alt+Shift to change the
keyboard to the Russian or Greek keyboard.
7.) Apply the changes by pressing Apply. Press OK TWICE
.
Figure 8-107 Advanced tab
NOTICE
To have the settings take effect, you MUST turn off the system and turn it back on.