Chapter 7: Digital Detector
5495975-1EN Rev.9 7-12
© 2013-2017 General Electric Company. All rights reserved.
Figure 7-16 Connect with detector
Detector Grid
The Digital Detector grid is integrated with a holder that fits the detector exactly. The grid fits over the
detector handle and has raised edges to fit around the detector. Once together, you may handle the grid
and detector as one unit.
The arrows on the grid surface show the direction of the grid lines, which run horizontally.
Figure 7-17 Detector Grid
Table 7-6 Detector Grid
The system will recognize when the grid is attached. Should the grid be attached to the backside of the
detector, the system will display an exposure inhibit.
Item Description
1
Grid (6:1 grid ratio horizontal is standard, 8:1 grid ratio horizontal is optional)
2
Detector
Front Side Back Side