Build Custom Looks
Looks are the way an image is processed to be viewed by a radiologist for interpretation.
Looks are pre-defined combinations of brightness, contrast, edge enhancement, and tissue equalization
(TE). Refer to Tissue Equalization Overview
(p. 15-55) for more information.
Factory Look descriptions:
Factory Look 1 – similar to analog film, low edge, no TE
Factory Look 2 – low CR look, low edge, moderate tissue contrast
Factory Look 3 – moderate CR look, low edge, TE
Factory Look 4 – highly digital look, high edge, high TE
Follow this process to create an new custom look.
Recommendation: When building a custom look, start with the factory look that is closest to the desired
result. Write down the values and apply them to a new custom process. Then change the applicable vari-
ables.
1. From the Worklist screen, click [UTILITIES].
2. Select Preferences > Image Processing.
3. Click Image Processing [EDIT].
4. The Image Processing Preference Editor (Figure 15-41) appears.
Figure 15-41 Image Processing Preference Editor
5. Select the Anatomy (for example: Chest, Abdomen, Lumbar Spine, etc.).
6. Select the View (for example: antero-posterior or lateral).