EasyManua.ls Logo

GE Optima XR646 - Position the Receptor; Vertical Adjustment

GE Optima XR646
463 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Chapter 8: System Hardware Overview
5495975-1EN Rev.9 8-46
© 2013-2017 General Electric Company. All rights reserved.
Position the Receptor
The wallstand receptor placement may be adjusted vertically.
Vertical Adjustment
The receptor arm assembly can be moved vertically (Figure 8-33) by the Vertical Adjust Handle manually.
Figure 8-33 Receptor and arm vertical movement
CAUTION Patients should be clear of the wallstand when vertical movement is in process.
CAUTION Be careful of the hand grips that stick out below the wallstand when positioning
wheelchair patients under the wallstand.

Table of Contents

Related product manuals