Chapter 8: System Hardware Overview
5495975-1EN Rev.9 8-12
© 2013-2017 General Electric Company. All rights reserved.
Overhead Tube Suspension (OTS)
The Overhead Tube Suspension (OTS) is the positioning device that supports the X-ray tube and OTS User
Interface. The OTS has available movement on 5 axes for use in advanced applications. The suspension
provides convenient movement and accurate positioning of the equipment.
Figure 8-6 Overhead Tube Suspension (OTS)
OTS User Interface
The OTS User Interface (Figure 8-7 and Table 8-6) allows you to make receptor, FOV, kV and mAs selec-
tions without returning to the Acquisition Workstation. This interface also provides the functions to move
the OTS in multiple directions.
Note: FOV (field of view) can only be changed on the OTS User Interface.
Note: Changing exposure parameters or receptors on the OTS User Interface or the Acquisition Work-
station will result in a change to both.
The X-ray OTS consists of the
following major elements:
1. Overhead stationary rail
and lateral bridge
2. Telescopic column and
carriage
3. X-Ray tube and housing
4. OTS User Interface
5. Multi-leaf collimator
(Refer to Chapter 8:
System Hardware
Overview-Multi-Leaf
Collimator (p. 8-26).)
2
4
3
1
5