EasyManua.ls Logo

GE Optima XR646 - Electro-Magnetic Interference

GE Optima XR646
463 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Chapter 7: Digital Detector
5495975-1EN Rev.9 7-20
© 2013-2017 General Electric Company. All rights reserved.
Figure 7-26 Detector fault condition
In this condition the battery LED’s may be lit to show the condition of the battery.
Electro-magnetic Interference
The detector has been designed and tested to meet all IEC regulations in regard to electro-magnetic (EM)
susceptibility (and EMC).
However, there are no regulations for low frequency EM fields. All flat panel detectors have some suscep-
tibility to these EM fields.
This detector is robust to EM fields up to 0.1mG over a broad low frequency range (DC - 250kHz).
It has been found that some common equipment in clinical environments can generate EM fields well
over 0.1mG. Some measurements have shown 4mG field strengths which can cause artifacts on x-ray
images.
EM fields are reduced by the square of the distance between the EM source and the detector. Thus, mov-
ing the EM source twice as far away will reduce the field strength by 4 times.
General suggestions:
Keep IV pumps, patient monitoring, feeding pumps 1 meter or more away from any detector surface.
Consider turning off equipment that cannot be moved.
Change the patient or detector orientation /position to maximize distance from any equipment.
Possible sources of EM fields:
IV pumps
Monitors
Feeding pumps
Patient monitors
ECG equipment

Table of Contents

Related product manuals