Chapter 8: System Hardware Overview
5495975-1EN Rev.9 8-35
© 2013-2017 General Electric Company. All rights reserved.
Position the Receptor
The wallstand receptor placement may be adjusted vertically or tilted.
Vertical Adjustment
The receptor arm assembly can be moved vertically (Figure 8-20) in one of three ways:
Using the Vertical Adjust and receptor Tilt Handle.
Using the foot control.
Using the positioning remote. Refer to Standard Table Hand Grips and Compression Band
(p. 8-72)for
more information.
Figure 8-20 Receptor and arm vertical movement
CAUTION Patients should be clear of the wallstand when vertical movement is in process.
CAUTION Be careful of the hand grips that stick out below the wallstand when positioning
wheelchair patients under the wallstand.