Chapter 14: Quality Assurance and Maintenance
5495975-1EN Rev.9 14-11
© 2013-2017 General Electric Company. All rights reserved.
Figure 14-13 QAP Results screen
Failed QAP
Note: In the event of a failed QAP test, repeat the QAP procedure to confirm the failure. This section
identifies common problems to check.
If a single failure is confirmed, call to schedule service. The system is operational, although inspection
and potential calibration are needed.
If multiple failures are confirmed, image quality may be effected; cease use of the unit and call for imme-
diate service.