Chapter 8: System Hardware Overview
5495975-1EN Rev.9 8-25
© 2013-2017 General Electric Company. All rights reserved.
a) Press and hold the Tube Angulation Lock Release button on the OTS user interface.
b) Move the tube unit to the desired angle.
c) Release the Tube Angulation Lock Release button.
2. Use the OTS Rotation Detent Release lever to rotate the tube about the vertical axis of the telescopic
column.
a) Press the Rotation Detent Release lever on the right side of the OTS.
b) Rotate the tube unit.
c) Release the lever. The OTS locks in the next 30 degree position.
CAUTION Potential Pinch Point: The area where the tube connects to the column may create
a pinch point when the tube is rotated. Operators should keep their hands on the
OTS handle and keep patient’s clear while rotating the tube.