Chapter 10: Image Acquisition
5495975-1EN Rev.9 10-4
© 2013-2017 General Electric Company. All rights reserved.
Receptor: Selects the receptor for the protocol.
The available receptors depend on your system
configuration and on the selected protocol. Not all
receptors are available for every protocol.
In order from left to right, the options are:
Wallstand
Table
Film/CR Cassette
Detector (Table Top Mode)
Note: The generator limits exposure time to two (2) sec-
onds for all receptors.
[AEC] and [FIXED] (mode) Selects AEC or FIXED modes.
Refer to Automatic Exposure Control (AEC)
(p. 10-24) for
more information about AEC.
Reset Technique: Resets the technique to the default protocol settings.
Grid status Shows the current grid status and the recommended grid
status for this technique. The alert icon indicates that the
grid is not in the recommended status. A checkmark
indicates that the current grid status matches the
recommended status.
Note: Exposure may not be inhibited if the current grid
status is not the recommended status.
SID Shows the current SID and the recommended SID for this
technique. The alert icon indicates that the SID is not at
the recommended status. A checkmark indicates that the
current SID matches the recommended SID.
Note: Exposure may not be inhibited if the current SID is
not the recommended SID.
Table 10-1 Image Acquisition functions
Function Description