Chapter 2: Safety and Regulatory 
5495975-1EN Rev.9 2-35
© 2013-2017 General Electric Company. All rights reserved.
2 Patient load label
WARNING
1. The table can be moved vertically when the load is less than or equal to 220 kg (485 lbs), 
located in the center of the tabletop end and the tabletop is positioned in the center.
2. The table can be moved vertically when the distributed load over the table is less than or 
equal to 320 kg (706 lbs) and the tabletop is positioned in the center.
3. The table cannot be moved vertically and the tabletop shall be positioned in the center 
when the distributed load over the table is more than 320 kg (706 lbs) and less than or 
equal to 400 kg (882 lbs). 
WARNING When the patient weight exceeds 220 kg (485 lbs) then patient on/off load of 
the table should only occur from the center of the front or rear side of the 
table. (Maximum weight 400 kg (882 lbs))
3 Inhibition warning label
Table 2-17 Digital Table warning Labels Icons 
Item Description Label