Chapter 4: General Information
5495975-1EN Rev.9 4-12
© 2013-2017 General Electric Company. All rights reserved.
CAUTION Ensure no detector is in X-Ray path
Follow this process to warm up the tube.
CAUTION X-rays are generated during tube warm up. Use proper radiation practices at all
times.
1. Check the room to make sure it is clear of other personnel before making X-ray exposures.
2. Click [WARM TUBE] on the Worklist screen.
Note: Because the tube warming procedure produces x-rays, it cannot be performed when an exam is
open. If the tube must be warmed before the exam can begin, suspend the exam and remove the
patient and any others from the room before warming the tube. Refer to Chapter 10: Image
Acquisition-End Exam (p. 10-31) for more information.
3. Correct any exposure holds.
CAUTION Exposures should not be made on any detector during tube warm-up. The system
will attempt to mis-align the tube and the detector, but you may need to manually
move the tube so that the detector is not exposed during the warm-up sequence.
Pay close attention to system message information for instructions to manually
move the detector. Exposing the detector may cause a “burned in” image artifact.
Table 4-1 Warm Tube functions
Function Description
Tube Heat Status Shows the current status of the tube.
Warm up needed in __ Min Shows how many minutes until the tube needs to be warmed.
A time of 0 minutes means that the tube must be warmed immediately.
Exposure Required Shows the number of exposures required to warm the tube.
Soak Time Remaining Shows how many seconds of soak time remain.
Exposure Hold Lists any inhibits to making an exposure or warming the tube.
kV Shows the kV of the current exposure.
mA Shows the mA of the current exposure.
mAs Shows the mAs of the current exposure.
mSec Shows the mSec of the current exposure.
[CLOSE] Closes the Warm Tube screen.