3-61
CHAPTER 3 Settings and Measurement
(1)-4 EU (Scaling settings)
This is used to convert the measured signals to other units.
(1) (1)
(
2
)
(3)
(3) (4)
(5) (6) (7)
<When input is voltage and effective value (RMS)> <When input is temperature>
Setting Description
(1) EU Set the scaling function to ON or OFF.
(2) Dec pt Set the decimal point position for an EU value.
(3) Meas. Value (Upper/Lower) Set the upper and lower limits of values to be converted.
* For temperature input, there is no distinction between upper and lower limits. See the
setting examples shown below for details.
(4) EU Value (Upper/Lower) Set the upper and lower limits of values to be converted.
* For temperature input, there is no distinction between upper and lower limits. See the
setting examples shown below for details.
(5) Select Select a specic engineering unit classication. (The following are available.)
Cur, Length, Area, Volume, Velocity, Accel, Freq, Mass, Energy, Pressure, Flow, Temp,
Strain, Brightness, Concentration; Optional
You can enter arbitrary units by selecting Arbitrary.
(6) Unit Select a unit to be used after conversion.
A unit displayed in this eld belongs to the classication selected in "Select".
To set a unit not displayed in this eld, set arbitrary text in "Arbitrary unit." The setting
selected in this eld is displayed in "Arbitrary unit." (Number of characters that can be
entered: Up to 8 characters)
(7) Reaeds the current
temperature
Substitute the currently measured value into (3) Meas. Value and (4) EU Value.
* The value is not substituted when burnout occurs or the scale is exceeded.
If a message window opens, follow the instruction in the message to change the setting value.
The Scaling function performs calculation using a ratio of the Meas. Value and EU Value settings.
When the digital display shows “++++/----”, the converted value will not be processed by this device.
The span may be changed depending on the Scaling settings.
For temperature input, the offset setting for an input value is used.
Dec pt
+5.000Upper limit
-5.000Lower limit
Meas. Value
+20.00
-20.00
EU Value
+ xx.xx
Unit
rpm
+5V
CH.1 10V
-5V
+ 20.00 rpm
CH.1 Scaling 1
- 20.00 rpm
Setting example: For voltage input
Meas. Value
3°C is always added to the measurement value.
22.0
EU Value
25.0
Setting example: For temperature input