3-102
CHAPTER 3 Settings and Measurement
Level and Edge operations
In the Level operation, a trigger is assumed to be generated if the trigger conditions are met when the [START]
key is pressed.
In the Edge operation, a trigger is not assumed to be generated even if the trigger conditions are met when
the [START] key is pressed.
A trigger is assumed to be generated when the trigger conditions, after not being met, are met again.
* A trigger is still assumed to be generated even if the trigger conditions are met once in the Edge operation
and then are no longer met.
If the START key is pressed here If the START key is pressed here
For 1CH, a trigger is assumed to
be generated here, too.
Trigger generation area
Example: Rising Example: Rising Edge And for 1CH and 2CH
Setting
level
An edge exceeding
the setting level is
required.
For 1CH, a trigger
is generated here.
For 2CH, a
starts
Measurement starts
from here in the
Edge operation.
Measrement is started here.
Measurement
immediately
starts in the
Level operation.
1CH
2CH
Trigger and Alarm operations (H (Rising, L (Falling), Window In, Window Out)
H (Rising) : A trigger/alarm is generated when the input signal reaches specied level.
Input signal
Measurement starts (alarm generated)
Trigger (alarm) generation area
Setting level
L (Falling) : A trigger/alarm is generated when the input signal reaches the specied level.
Input signal
Measurement starts (alarm generated)
Trigger (alarm) generation area
Setting level
Win In : Used to specify the upper and lower limits. When the input signal level is enters specied limit
levels, a trigger/alarm is generated.
Input signal
Measurement starts (alarm generated)
Trigger (alarm) generation area
Upper limit level
Lower limit level
Win Out : Used to specify the upper and lower limits. When the input signal level exits specied limit levels, a
trigger/alarm is generated.
Input signal
Measurement starts (alarm generated)
Trigger (alarm) generation area
Upper limit level