EasyManua.ls Logo

GRAPHTEC GL980 - Trigger Functions; Alarm Functions

GRAPHTEC GL980
202 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
4-8
CHAPTER 4 Specication
Trigger Functions
Item Description
Repeat Trigger Off, On
Repeat interval Start to start interval, stop to start interval
1 second - 9999 hours 59 minutes 59 seconds
Trigger types Start: Data capture starts when a trigger is generated.
Stop: Data capture stops when a trigger is generated.
Trigger conditions Start: Off, level, alarm, external input, specied time, specied day of the week, certain
time, hourly
Stop: Off, level, alarm, external input, specied time, specied day of the week, certain time
* Level can be set to each CH.
Trigger judgment modes Analog judgment: H/Rising, L/Falling, Window In, Window Out
Logic judgment: H, L (Pattern comparison)
Pulse judgment: H/Rising, L/Falling, Window In, Window Out
Combination of channel Level OR, Level AND, Edge OR, Edge AND
Alarm Functions
Item Description
Alarm judgment modes Analog judgment: H, L, Window In, Window Out
Logic judgment: H, L
Pulse judgment: H, L, Window In, Window Out
Alarm detection cycle Linked to the sampling interval (Fastest 1 μs)
Alarm generation hold function ON, OFF Setting available
Combination of channel Level OR xed