3-66
CHAPTER 3 Settings and Measurement
Logic and Pulse settings
For settings related to digital input.
< Pulse setting > < Logic setting >
(1)-6 Logic/Pulse
To select the processing method for digital input.
Selection item Description
Off Digital input measurement is disabled.
Logic Digital input is processed as logic signals.
Pulse Digital input is processed as pulse signals.
* When pulse input is selected, the sampling interval for pulse is displayed. For details of sampling intervals
for pulse, refer to "3.5 Setting Menus" - "(2) DATA setting" - " Sampling interval".
(1)-7 Input
To set the pulse measurement mode. This setting is available only if Pulse is selected in (1)-6.
Selection item Description
Off Pulse input measurement is disabled.
Revol. The number of pulses per sample interval is counted and converted to the number of revolutions per
minute.
Counts Captures the cumulative number of pulses for each sampling interval from the start of measurement.
Inst. Captures the number of pulses for each sampling interval.
(1)-8 Filter
To set the lter for input.
Selection item Description
Off Disables hardware lter.
On Enables hardware lter. It is effective in a noisy environment.
The lter is approximately 30 Hz (-3 dB).