1-3
CHAPTER 1 General Description
1.3 Operating Environment
This section explains the operating environment for this device.
Ambient Operating Conditions
(1) Ambient temperature and humidity (Use within the following range.)
• Temperature range: 0 to 40°C (When AC adapter or battery is operated)
: 15 to 25°C (Battery being charged)
• Humidity range: 5 to 85% R.H. (non-condensing)
(2) Environment (Do not use in the following conditions.)
• A Location exposed to direct sunlight
• Locations exposed to salty air, corrosive gases, or organic solvents
• Dusty locations
• Locations subject to vibration or impact
• Locations subject to voltage surge or electromagnetic interference such as lightning or electric furnaces
(3) Installation category (over-voltage category)
• This device belongs to Installation Category II dened in IEC60664-1.
• Never use this device for Installation Category III or IV.
(4) Measurement category
• This device is not available for Measurement category II, III and IV.
If condensation occurs...
Condensation occurs in the form of water droplets on the device surfaces and interior when this device
is moved from a cold to a warm location. Using this device with condensation will cause malfunctioning.
Wait until the condensation has disappeared before turning the power on.
Warming-up Before Use
This device should be allowed to warm up with the power turned on for approximately 30 minutes to ensure
that it operates according to the specied performance.