3-121
CHAPTER 3 Settings and Measurement
During data replaying
(1)
(2)
(3)
(4)
(1)
(2)
(3)
(4)
(5)
(6)
Setting Description
(1) File Type Set the le format of data.
GBD: The data le is created in our Graphtec's binary format.
* Data cannot be changed.
CSV: The data le is created in text format.
* While GBD le is being replayed, you can duplicate it to a GBD le and convert it to a CSV
le.
* While CSV le is being replayed, you can copy to CSV le, but conversion to GBD le
cannot be performed.
(2) Name Type Set the naming method for the data le.
Auto: File name is appended automatically.
Example) 20170201-123456.GBD
Numbers part ..... Date and time when the le was created
* Example: 2017 year 2 month 1 day 12 hour 34 minute 56 second
GBD .................... Data format
GBD (binary data)
CSV (text format)
Arbitrary: Data is captured with the le name you entered.
Serial number: File is created by appending a Serial number to the arbitrarily entered le
name.
[Auto] (3) Folder Set the folder of the save destination.
[Arbitrary]
[Serial
number]
(3) Folder
File
Set the data le name.
The folder for saving the le is displayed in the Folder.
(4) Save Range Set the range of data to be saved.
All Data: All the data is saved regardless of the cursor.
Data between cursors: Data of the range between A and B cursors is saved.
(5) CSV decimal point Set the decimal point character of the CSV le.
Period (.), Comma (,)
(6) CSV delimiter Set the delimiter of the CSV le.
Comma (,), Tab, Semicolon (;)
(7)-3 Immediate backup
Perform backup processing immediately.
During the recording, you can use this function when backup setting is set to On.
Select the Immediate Backup and press [ENTER] key to perform the backup process immediately regardless
of the backup interval set in the backup setting.
* Back up data that has not been backed up yet is backed up.
* If the amount of data to be backed up increases by combining the sampling interval and backup interval, it
may take time to perform the backup process.