3-120
CHAPTER 3 Settings and Measurement
(7)-2 Data Save
Data in the internal RAM can be saved as well.
During data replay, you can save the data that is being replayed.
When saving in internal RAM
(1)
(2)
(3)
(4)
(5)
(1)
(2)
(3)
(4)
(5)
(6)
(7)
Setting Description
(1) Save block Select the internal RAM block to be saved.
1 to 8, ALL
Numbers vary depending on the number of block divisions.
When ALL is selected, the block in which all the data exists is saved.
(2) Blocks status The status of the internal RAM is displayed.
(3) File Type Set the le format of data.
GBD: The data le is created in our own binary format.
* Data cannot be changed.
CSV: The data le is created in text format.
(4) Name Type Set the naming method for the data le.
Auto: File name is appended automatically.
Example) 20170201-123456.GBD
Numbers part ..... Date and time when the le was created
* Example: 2017 year 2 month 1 day 12 hour 34 minute 56 second
GBD .................... Data format
GBD (binary data)
CSV (text format)
Arbitrary: Data is captured with the le name you entered.
Serial number: File is created by indexing a serial number to the arbitrarily entered le name.
[Auto] (5) Folder Set the folder of the save destination.
[Arbitrary]
[Serial
number]
(5) Folder
File
Set the le name of the data.
The folder for saving the le is displayed in the Folder.
(6) CSV decimal point Set the decimal point character of the CSV le.
Period (.), Comma (,)
(7) CSV delimiter Set the delimiter of the CSV le.
Comma (,), Tab (Tab), Semicolon (;)