EasyManua.ls Logo

GRAPHTEC GL980 - Trigger Level Settings;Alarm Level Settings

GRAPHTEC GL980
202 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
3-101
CHAPTER 3 Settings and Measurement
Trigger level settings/Alarm level settings
Species detailed conditions for each channel when the start and stop side source settings are Level.
The conguration of the level trigger is as shown in the gure below.
* Pulse and Logic are switchable.
CH n
Mode Level
Combination
Pulse n
Mode Level Trigger
Logic n
Mode
Lower-Level-Upper
* Pulse and Logic are switchable.
* Specify an alarm output destination for each channel and Pulse/Logic.
Each of the alarms is ORed at the output destination.
Example: If you specify 1 as the output destination of 1CH and 2CH and 2 as that of 3CH
and 4CH, Alarm Output 1 occurs when one of 1CH and 2CH meets the conditions,
and Alarm Output 2 occurs when one of 3Ch and 4CH meets the conditions.
CH n
Mode Level
OR
Pulse n
Mode Level Alarm output n
Logic n
Mode Level
< Trigger setting > < Alarm setting >
(2) (2)(3) (3) (4)
(1)
Setting Description
(1) Combination Set the combination of the trigger condition set for each channel.
Level OR: Starts (stops) capturing when a set trigger condition is satised.
The level operation is determined by each condition.
Level AND: Starts (stops) capturing when all the set trigger conditions are satised.
The level operation is determined by each condition.
Edge OR: Starts (stops) capturing when a set trigger condition is satised.
The each operation is determined by each condition.
Edge AND: Starts (stops) capturing when all the set trigger conditions are satised.
The each operation is determined by each condition.
* Only for trigger setting.
(2) Mode Set the mode to compare trigger/alarm for each channel.
Off: Trigger/alarm of set CH is disabled.
H (Rising): Trigger/alarm is established when the input signal exceeds the set level.
L (Falling): Trigger/alarm is established when the input signal is lower than the set level.
Window In: Trigger/alarm is established when the input signal is within the range between
the upper level and lower level.
* Logic CH has no setting.
Window Out: Trigger/alarm is established when the input signal is out of the range between
the upper level and lower level.
* Logic CH has no setting.
(3) Level Set the level to compare trigger/alarm.
Set the comparison level to one place when the mode is H (Rising) or L (Falling).
Set the comparison level to two places when the mode is Window In or Window Out.
(4) Output Set the alarm output CH when an alarm occurs.
* Only for alarm setting
* The operation in parentheses () is for edge operation.
* When the alarm is set, the level operation is xed.