3-58
CHAPTER 3 Settings and Measurement
Analog settings
Settings the conditions for analog signals.
When you use CH ALL to set input, range, lter and EU, all channels are set to the same values when input
is the same. Range is set only for the same input channels. However, the range of a channel is not changed
if EU (scaling) is set to On.
CH ALL - Misc. - Span Settings is set only for the same range channels.
(1)-1 Input
Select input conditions.
Selection item Description
Off Input signal measurement is disabled. No waveform or digital value is displayed.
DC Used for measuring direct-current voltage.
RMS Used to measure the RMS of AC voltage.
TEMP Used for measuring temperature.
RH Used for measuring humidity with the humidity sensor B-530.
In this case, the voltage range will become 1 V, and the EU settings will not be available
* Enter DC-RMS as the effective value (RMS).
(1)-2 Range
Select the range to measure.
Input item Description
DC 20, 50, 100, 200, 500 mV, 1, 2, 5, 10, 20, 50, 100, 200, 500 V, 1-5 V
RMS 10, 25, 50, 100, 250, 500 mVrms
1, 2.5, 5, 10, 25, 50, 100, 250 Vrms
TEMP TC-K, TC-J, TC-T, TC-R, TC-E, TC-B, TC-S, TC-N, TC-W
RH 100% xed