2-22
CHAPTER 2 Checks and Preparation
During data capture, when the battery power is at low, the capturing stops automatically.
When using in conjunction with AC adapter, this device automatically shifts to battery-powered at the time
of power outage.
At such time, pay attention to the remaining battery power. If the low-charged battery is inserted, this
device may stop abnormally due to power failure or instantaneous interruption.
While power is supplied directly from the DC power without the AC adapter link, and the DC voltage
reaches 16V or less, the battery charge disables.
When an empty battery is charged in other recorder GL220/820/900 units, the charging will stop at about
80-90%. Disconnect and re-connect the AC adapter, or remove and insert the battery pack to start charging
the battery to reach up to 100%. (depending on the remaining capacity.)