6.3 Measurement examples
99
6
Chapter 6 Measurements
*
only when selecting a network B1 or B2
1. Connect the instrument and the equipment under test.
See "3.3 Connecting equipment under test to the instrument" (p. 36)
2. Connect test leads to terminals T1 and T2.
3. Position the tips of test leads on the applied part of the equipment under test.
4. During manual measurement, patient auxiliary current measurement starts upon return to the
measurement screen.
Measurement conditions can be changed during measurement.
During automatic measurements, becomes available upon return to the measurement
screen. Press to start patient auxiliary current measurement.
Only is available during automatic measurements.
Patient auxiliary current measurement
*
The terminal block is not used for internally powered equipment.
1
2
3
Terminal T1 and one terminal
on applied part
Terminal T2 and another
terminal on applied part
4