Appendix2 List of instrument status, other test condition and special test condition
A4
When network A, D, E or F is selected
When network B1 is selected
Appendix2 List of instrument status, other test
condition and special test condition
Measurement
mode
Class-I equipment Class-II equipment Internally powered equipment
Equipment status Equipment status Equipment status
Normal
condi-
tion
Single-fault condition
Normal
condi-
tion
Single-fault condition
Normal
condi-
tion
Single-fault condition
Power
cord dis-
connec-
tion
Ground-
ing wire
discon-
nection
Applica-
tion of
line volt-
age L
Applica-
tion of
line volt-
age N
Power
cord dis-
connec-
tion
Ground-
ing wire
discon-
nection
Applica-
tion of
line volt-
age L
Applica-
tion of
line volt-
age N
Power
cord dis-
connec-
tion
Ground-
ing wire
discon-
nection
Applica-
tion of
line volt-
age L
Applica-
tion of
line volt-
age N
Earth leakage
current
–––––––––––––
Enclosure -
Line leakage
current
––––––––– ––
Enclosure -
Earth leakage
current
–– ––– ––––
Enclosure -
Enclosure
current
–– ––– ––––
Free current*
––
––– ––––
*Only network A
:Can be set, –:Cannot be set
Measure-
ment
mode
Class-I equipment Class-II equipment Internally powered equipment
Equipment status Equipment status Equipment status
Normal
condition
Single-fault condition
Normal
condition
Single-fault condition
Normal
condition
Single-fault condition
Power
cord dis-
connec-
tion
Ground-
ing wire
discon-
nection
Applica-
tion of
110%
voltage
: Positive
phase
Applica-
tion of
110%
voltage
: Nega-
tive
phase
Power
cord dis-
connec-
tion
Ground-
ing wire
discon-
nection
Applica-
tion of
110%
voltage
: Positive
phase
Applica-
tion of
110%
voltage
: Nega-
tive
phase
Power
cord dis-
connec-
tion
Ground-
ing wire
discon-
nection
Applica-
tion of
110%
voltage
: Positive
phase
Applica-
tion of
110%
voltage
:Nega-
tive
phase
Earth
leakage
current
––– – –––– – ––––
Enclosure
- Earth
leakage
current
–
––
Enclosure
- Enclo-
sure
current
–
––
Patient
auxiliary
current
––
–––
––––
Patient
leakage
current I
––
–––
––––
Patient
leakage
current
II(only
B-type ap-
plied parts)
–––
–––
–––
Patient
leakage
current
III(only BF-
type or CF-
type ap-
plied parts)
–––
–––
–––
Free
current
–
––
:Can be set, –:Cannot be set