Appendix3 List of default settings
A7
Appendix
Default settings for each network when the software is initialized.
Network A
After selecting network, class and leakage current mode
Appendix3 List of default settings
default setting
Leakage current mode item Class I Class II Internally powered
Earth leakage current
measurement manual
––
polarity positive phase
equipment status normal
filter ON
target current ACDC
range AUTO
allowable value : upper
limit : normal
1.000 mA
allowable value :
upper limit : fault
1.000 mA
allowable value : lower
limit : ON/OFF : normal
OFF
allowable value : lower
limit : ON/OFF : fault
OFF
Enclosure - Earth
leakage current
measurement manual manual
polarity positive phase positive phase
equipment status normal normal
filter ON ON ON
target current ACDC ACDC ACDC
range AUTO AUTO AUTO
allowable value : upper
limit : normal
1.000 mA 1.000 mA 1.000 mA
allowable value : upper
limit : fault
1.000 mA 1.000 mA
allowable value : lower
limit : ON/OFF : normal
OFF OFF OFF
allowable value : lower
limit : ON/OFF : fault
OFF OFF
Enclosure - Enclosure
leakage current
measurement manual manual
polarity positive phase positive phase
equipment status normal normal
filter ON ON ON
target current ACDC ACDC ACDC
range AUTO AUTO AUTO
allowable value :
upper limit : normal
1.000 mA 1.000 mA 1.000 mA
allowable value :
upper limit : fault
1.000 mA 1.000 mA
allowable value :
lower limit : ON/OFF
: normal
OFF OFF OFF
allowable value : lower
limit : ON/OFF : fault
OFF OFF