8.7 Command Message Description
165
8
Chapter 8 Controlling the Instrument From a Computer
Specific commands
Queries the automatic measurement completion
Syntax
Query
Response
:AMC?
<numerical value>
<numerical value> = 0/1 (NR1)
0: In automatic measurement
1:
Automatic measurement completed
Function
Query Returns automatic measurement condition as a numeric value.
Example
Query
Response
:AMC?
:AMC 1 (Headers: ON)
1 (Headers: OFF)
Automatic measurement has been completed.
Error
An execution error occurs in the following cases :
When the measurement method is set to "manual measurement"
When the measurement mode is not selected
Note
Use this command after executing the :MODE command to activate the mea-
surement mode.
Sets and queries the voltage application ON/OFF
Syntax
Command
Query
Response
:APPLy <characters>
:APPLy?
<characters>
<characters> = ON/OFF
ON: Starts applying voltage.
OFF: Stops applying voltage.
Function
Command When network A, C, D, E, or F is selected
In manual measurement, the application of line voltage starts.
When network B1 or B2 is selected
In manual measurement, the output of an application of 110% voltage starts.
Single-fault condition(application of 110% of rated voltage), Other test condi-
tion (application of 110% of rated voltage), Special test condition(application
of 110% of rated voltage)
Query Returns the voltage application condition as alphabetic characters.
Example
Command
:APPLy ON
Starts the output of an application of 110% voltage.
(When network B1 is selected)
Query
Response
:APPLy?
:APPLY ON (Headers: ON)
ON (Headers: OFF)
The voltage application status has been set.