7.8 Auto Measure Setting (set instrument status)
129
89
7
Chapter 7 Instrument System Settings
When setting “Last Measurement State”
When setting “Normal State”
7.8 Auto Measure Setting
(set instrument status)
When setting Auto Measure Setting is "Last Measurement
State"(default), the instrument stores the final state (fault
condition etc.).
Example:
If disconnection of one wire in the power cord is the final
test item of the automatic measurement, no power is sup-
plied to the device under test after the completion of an
automatic measurement.
Screen for Auto Measurement Setting
Screen for Auto Measurement Setting
When Auto Measure Setting is "Normal State", the final
state (fault condition, etc.) in an automatic measurement is
not stored and the device under test is automatically sup-
plied with power. (It is returned to normal condition)