7.2 Panel Load(Loading Saved Measurement Conditions)
120
Measurement conditions saved by using the measurement screen can be loaded (read).
The following lists the saved measurement conditions.
7.2 Panel Load(Loading Saved Measure-
ment Conditions)
Measurement items/target current mode/network/measurement range/filter setting/
allowable value (normal condition, fault condition)/fault condition setting/power supply polarity/
measurement class/applied part condition/equipment name/control number/measurement
items for automatic measurement/measuring time/measurement delay time
1. Press on the initial screen to display the system screen.
2. Press to display the panel load screen.
The screen shows the keys of saved panel numbers. The protruding black keys can be selected.
Press to go between pages.
3. Press a key showing the panel number
to load. A confirmation screen appears.
4. Press to load data.
The read measurement conditions are
shown on the screen.
Press to return to the previous
screen.
Panel load screen
Panel load confirmation screen
Measurement condition data is saved in the instrument's internal SRAM.
The lithium battery provides power for memory backup. When battery power
becomes depleted, the saved measurement condition data in memory will be
lost.
If this happens, contact our service center for a battery replacement (at a
charge). The average service life of the lithium battery is about ten years.
When the conditions for leakage current between enclosure and line, patient
leakage current II, or patient leakage current III measurements are included,
measurement does not have allowable values in normal condition. Therefore,
the confirmation screen does not display
"UPPER_NORM : (value)."
"LOWER_NORM : (value)."