10.6 System-related Functions
263
8
9
Chapter 10 Specifications
10
10.6 System-related Functions
Beep sound setting (a) Allowable value judgment :
Selectable from "ON when Pass," "ON when FAIL," and "OFF"
(b) Key input : Selectable from ON/OFF
(c) T3 (110% voltage application terminal) output : Selectable from ON/OFF
*
(d) Line voltage output from T2 : Selectable from ON/OFF
Save/load function 30 panels for saving the following setting data
(measurement mode, network, equipment name, control number, grounding
class, applied part, measurement range, filter, target current, allowable value
setting, fault condition setting, power supply polarity switching, automatic
measurement items, automatic measuring time, measurement delay time,
Panel name)
Data save function Saved content : Equipment under test information (equipment name,
serial number), measurement data, date
Memory capacity : Data on up to 100 units
Clock function Auto calendar, automatic leap-year adjustment, 24 hour clock
Clock accuracy: Deviation of about 4 minutes a month
Data backup function SRAM (Set condition), RTC
Backup battery life: Approx. 10 years (reference value at 25
°C (77 °F))
Backlight automatic OFF
function
(a) Constant ON (default)
(b) Auto OFF (1 min. to 30 min., in 1 min. increments)
The backlight automatically turns off after the set time elapses, turns ON
when a key is pressed, then turns OFF again after the set time elapses.
Self-test function (a) MEM (internal RAM)
(b) KEY (6 x 6 matrix touch panel)
(c) LCD (front LCD panel)
(d) LED (warning lamp, LCD backlight)
(e) Beep sound
Language setting (a) Japanese
(b) English (default)
Fuse blown check function Checks for blown fuses in the network
At power on (when set) or key operation in the system screen
Relay check function Checks whether the network relay works normally
At power on (when set) or key operation in the system screen
Allowable lower limit value
setting
Sets a lower limit value for all measurements.
System reset (a) Returns instrument to its factory defaults
(includes measurement data, setting conditions, etc.)
(b) Clears all measurement data.
(c) Clears all condition setting data including panels.
*
only ST5540