7.6 Self-Test
125
89
7
Chapter 7 Instrument System Settings
When is selected
The black and white sections alternate on the screen.
Check the screen for any malfunctioning dots.
If there are any malfunctioning dots, the screen is not
functioning normally and you should contact your
dealer or Hioki representative.
After the check, touch any location on the screen to
exit test mode and return to the self-test screen.
When is selected
Press to activate the buzzer.
Press to stop the buzzer.
If the buzzer does not sound, the instrument requires
repair. Contact your dealer or Hioki representative.
Press to return to the self-test screen.
When is selected
The DANGER lamp and the LCD backlight flash
alternately. Confirm that all the lamps operate.
If any lamp does not turn on, the instrument requires
repair. Contact your dealer or Hioki representative.
Press to return to the self-test screen.