8.7 Command Message Description
182
Patient leakage
current (SIP/SOP)
Allowable value data for normal
condition(upper limit)
Set value Set value
Set value
Allowable value for fault
condition(upper limit)
zero
Patient leakage
current
(F-type applied part)
Allowable value data for normal
condition(upper limit)
–
zero
–
zero
–
zero
Allowable value for fault
condition(upper limit)
Set value Set value Set value
Patient leakage
current
(Metal accessible part)
Allowable value data for normal
condition(upper limit)
zero
–
zero
–
zero
–
Allowable value for fault
condition(upper limit)
Set value Set value Set value
Total patient leakage
current
(Patient - Earth)
Allowable value data for normal
condition(upper limit)
Set value Set value
Set value
Allowable value for fault
condition(upper limit)
zero
Total patient leakage
current
(SIP/SOP)
Allowable value data for normal
condition(upper limit)
Set value Set value
Set value
Allowable value for fault
condition(upper limit)
zero
Total patient leakage
current
(F-type applied part)
Allowable value data for normal
condition(upper limit)
–
zero
–
zero
–
zero
Allowable value for fault
condition(upper limit)
Set value Set value Set value
Total patient leakage
current
(Metal accessible part)
Allowable value data for normal
condition(upper limit)
zero
–
zero
–
zero
–
Allowable value for fault
condition(upper limit)
Set value Set value Set value
Patient auxiliary current
Allowable value data for normal
condition(upper limit)
Set value Set value
Set value
Allowable value for fault
condition(upper limit)
zero
Free current
Allowable value data for normal
condition(upper limit)
Set value Set value
Set value
Allowable value for fault
condition(upper limit)
zero
Leakage current mode Item
Class I
Class II
Internally
powered
Type
B
Type
BF
Type
CF
Type
B
Type
BF
Type
CF
Type
B
Type
BF
Type