8.7 Command Message Description
204
Function
Command Sets the wait time for switching operation in automatic measurement (delay c).
Query Returns the wait time setting (delay c) for switching operation in automatic mea-
surement as a numeric value.
Example
Command
:CONFigure:AUTO ON;:CONFigure:WTIMe:ETC 10
Sets the wait time (delay c) for switching operation in automatic measurement
to 10 seconds.
Query
Response
:CONFigure:AUTO ON;:CONFigure:WTIMe:ETC?
:CONFIGURE:WTIME:ETC 1200 (Headers: ON)
1200 (Headers: OFF)
The wait time (delay c) for switching operation in automatic measurement has
been set to 1,200 seconds.
Error
An execution error occurs in the following cases:
• When the measurement method is set to "manual measurement"
• When the measurement mode is not selected
• When a mode other than "leakage current mode" is set
• In automatic measurement(Command)
• When a value outside the specified range is input
Note
• Use this command after executing the :MODE command to activate the mea-
surement mode.
• The unit is [s]. Fractions are rounded up when 0.5 or greater and rounded off
when less than 0.5.
Sets and queries the wait time for switching operation in automatic measure-
ment (delay c)
Sets and queries the wait time for switching operation in automatic measure-
ment (delay a)
Syntax
Command
Query
Response
:CONFigure:WTIMe:LINE <numerical value>
:CONFigure:WTIMe:LINE?
<numerical value>
<numerical value> = 0 to 1800 (NR1)
Function
Command Sets the wait time (delay a) for switching operation in automatic measurement.
Query Returns the wait time setting (delay a) for switching operation in automatic
measurement as a numeric value.
Example
Command
:CONFigure:AUTO ON;:CONFigure:WTIMe:LINE 10
Sets the wait time (delay a) for switching operation in automatic measurement
to 10 seconds.
Query
Response
:CONFigure:AUTO ON;:CONFigure:WTIMe:LINE?
:CONFIGURE:WTIME:LINE 1200 (Headers:ON)
1200 (Headers:OFF)
The wait time (delay a) for switching operation in automatic measurement has
been set to 1,200 seconds.
Error
An execution error occurs in the following cases :
• When the measurement method is set to "manual measurement"
• When the measurement mode is not selected
• When a mode other than "leakage current mode" is set
• In automatic measurement(Command)
• When a value outside the specified range is input