8.7 Command Message Description
214
Error
If the measurement mode is not set, an execution error occurs.
When there is no maximum value, +9.999E+10,3,0,0,0,0,0 will be displayed.
Note
To obtain the maximum value acquired in automatic measurement, execute
this command after confirming the completion of automatic measurement
(
:AMC? command). If the command is executed during automatic measure-
ment, the maximum value at command execution is obtained.
Use this command after executing the
:MODE command to activate the mea-
surement mode.
Queries the maximum value
Maximum
value
Judge
ment
Power
supply
polarity
Equipment under test status
Target
current
Other 110%
voltage
application
Specific
110% volt-
age applica-
tion
2.345 mA FAIL
Negative
phase
Single-fault condition(Disconnec-
tion of protective earth conductor)
AC+DC none none
Queries the protective conductor current measurement value
Syntax
Query
Response
:MEASure:PCC?
<numerical value>
<numerical value>=Measurement value(NR3)
Function
Example
Query Returns protective conductor current measurement value as a numeric value.
(unit : A)
Query
:MEASure:PCC?
Error
Response
:MEASURE:PCC +2.345E-03 (Headers : ON)
+2.345E-03 (Headers : OFF)
Protective conductor current measurement value is 2.345 mA.
Mode is set to something other than "protective conductor ammeter mode"
Note
Use this command after using the :SYSTem:MODE command to set the
protective conductor ammeter mode.
Queries the voltage measurement value
Syntax
Query
Response
:MEASure:VOLTage?
<numerical value>
<numerical value> = Measurement value (NR3)
Function
Query Returns the voltage measurement value as numeric values.(unit : V)
Example
Query
Response
:MEASure:VOLTage?
:MEASURE:VOLTAGE +2.345E+00 (Headers : ON)
+2.345E+00 (Headers : OFF)
The voltage measurement value is 2.345 V
Error
An execution error occurs when mode is set to something other than "voltmeter
mode."