Appendix3 List of default settings
A9
Appendix
Network B1
After selecting network, class, applied part and leakage current mode
When network B1 is selected, measured AC+DC current does not appear on the
screen. AC+DC is automatically selected internally.
default setting
Class I Class II Internally powered
Leakage
current
mode
item Type B Type BF Type CF Type B Type BF Type CF Type B Type BF Type CF
Earth leak-
age current
measurement manual
––––––
polarity positive phase
equipment
status
normal
s10 OFF
s12 OFF
s13
filter ON
target current ACDC
range AUTO
allowable value :
upper limit :
normal
500.0
A
allowable value :
upper limit : fault
1.000 mA
allowable value :
lower limit : ON/
OFF : normal
OFF
allowable value :
lower limit : ON/
OFF : fault
OFF
Enclosure -
Earth leak-
age current
measurement manual manual manual
polarity positive phase positive phase
equipment sta-
tus
normal normal normal
s10 OFF OFF
s12 OFF OFF
s13
filter ON ON ON
target current ACDC ACDC ACDC
rangeAUTOAUTOAUTO
allowable value :
upper limit :
normal
100.0
A 100.0 A 100.0 A
allowable value :
upper limit : fault
500.0
A 500.0 A 500.0 A
allowable value :
lower limit : ON/
OFF : normal
OFF OFF OFF
allowable value :
lower limit : ON/
OFF : fault
OFF OFF OFF
*
( ) is the value after DC switching