Appendix3 List of default settings
A11
Appendix
Patient leak-
age current
II
measurement manual
––
manual
––
manual
––
polarity positive
phase
positive
phase
–
equipment sta-
tus
110%
applied
N
110%
applied
N
110%
applied
N
s10 OFF OFF –
s12–––
s13 OFF OFF –
filter ON ON ON
target current ACDC ACDC ACDC
range AUTO AUTO AUTO
allowable value :
upper limit :
normal
–––
allowable value :
upper limit : fault
5.000
mA
5.000
mA
5.000
mA
allowable value :
lower limit : ON/
OFF : normal
–––
allowable value :
lower limit : ON/
OFF : fault
OFF OFF OFF
Patient leak-
age current
III
measurement
–
manual
–
manual
–
manual
polarity positive phase positive phase –
equipment
status
110%
applied N
110%
applied N
110%
applied N
s10 OFF OFF –
s12–––
s13 OFF OFF –
filter ON ON ON
target current ACDC ACDC ACDC
range AUTO AUTO AUTO
allowable value :
upper limit :
normal
–––
allowable value :
upper limit : fault
5.000
mA
50.00
A
5.000
mA
50.00
A
5.000
mA
50.00
A
allowable value :
lower limit : ON/
OFF : normal
–––
allowable value :
lower limit : ON/
OFF : fault
OFF OFF OFF
default setting
Class I Class II Internally powered
Leakage
current
mode
item Type B Type BF Type CF Type B Type BF Type CF Type B Type BF Type CF
*
( ) is the value after DC switching