Appendix3 List of default settings
A19
Appendix
Total patient
leakage cur-
rent (Patient
- Earth)
measurement manual manual manual
polarity positive phase positive phase –
equipment sta-
tus
normal normal normal
Other applied
voltages
–––
Application of
specific voltage
–––
measurement
terminals
– – T2T1
s10 OFF OFF –
s12–––
s13 OFF OFF –
filter ON ON ON
target current AC AC AC
range AUTO AUTO AUTO
allowable value
: upper limit :
normal
*
500.0 A
(50.00
A)
50.00
A
(50.00
A)
500.0
A
(50.00
A)
50.00
A
(50.00
A)
500.0
A
(50.00
A)
50.00
A
(50.00
A)
allowable value
: upper limit :
fault
*
1.000 mA
(100.0 A)
100.0
A
(100.0
A)
1.000 mA
(100.0
A)
100.0
A
(100.0
A)
–
allowable value
: lower limit :
ON/OFF :
normal
OFF OFF OFF
allowable value
: lower limit :
ON/OFF : fault
OFF OFF –
default setting
Class I Class II Internally powered
Leakage
current mode
item Type B Type BF Type CF Type B Type BF Type CF Type B Type BF Type CF
*