144
Input/Output/Measurement Specications
Specications of Model Z2011 Humidity Sensor
Temperature
measurement
accuracy
±0.5°C (10°C to 60°C)
Measuring temperature exceeding the range enclosed in parentheses above reduces this
accuracy as follows:
Add 0.015°C/°C (from −40°C inclusive to 10°C exclusive)
Add 0.02°C/°C (from 60°C exclusive to 80°C inclusive)
Humidity
measurement
accuracy
±3% RH (20°C to 30°C, 20% RH to 90% RH)
Measuring humidity exceeding the range enclosed in parentheses above reduces this
accuracy as the following table.
Values marked with an asterisk (*) are out of the guaranteed accuracy range (values used as reference).
800 10
20
40
60
20 30
80
100
0
40 50 60 70
±8% RH*
±8% RH*±8% RH*
±8% RH*
±8% RH
±6% RH
±6% RH
±5% RH
±5% RH
±5% RH
±5% RH
±5% RH
±4% RH
±4% RH
±6% RH
±6% RH
±6% RH*
±6% RH*
±10% RH*
±12% RH*
±12% RH*
±12% RH*±6% RH*
±3% RH
±3.5% RH
Temperature [°C]
Relative humidity [% RH]
Hysteresis: ±1% RH
Response time Approx. 300 s
(Temperature: 90 percentile response time during a temperature change from 0°C to 60°C
or from 60°C to 0°C [used as reference])
(Humidity: 90 percentile response time during a humidity change from 10% RH to 80% RH
or from 80% RH to 10% RH [used as reference])